3rd Generation Fluorescent Dye Technology, 10 Part Hematology, 88 Parameters, 100 sample per hour, that uses third-generation fluorescence flow cytometry and impedance technologies. It delivers full CBC with 6-part differential and built-in NRBC, RET, LW, BF, PLT-O, PLT-F, Abnormal WBC, Automatic rerun function, without the need for additional reagents. With high precision, multiple sample modes, and automation-ready features like a 50-tube autoloader, the F80 is designed for medium to high-volume clinical laboratories.
The Maccura F880 combines multiple advanced technologies to provide precise and comprehensive hematology results in a single run. Utilizing fluorescence-based nucleic acid staining and multi-angle flow cytometry alongside impedance and colorimetric methods, it achieves detailed 8-part white blood cell differentiation and accurate measurement of other critical parameters including reticulocytes, NRBCs, and platelets.
Designed to handle diverse clinical samples—whole blood, capillary blood, prediluted samples, and body fluids—the F880 ensures versatility and flexibility in laboratory workflows. Its built-in 50-tube autoloader with barcode reading streamlines sample processing, reduces manual workload, and increases throughput.
The analyzer delivers up to 37 reportable parameters with advanced flagging systems that detect abnormal cell populations and morphological changes, assisting early diagnosis of hematologic diseases such as infections, leukemia, and anemia. The intuitive touchscreen interface and bi-directional LIS connectivity facilitate user-friendly operation and seamless integration into laboratory information systems.
Overall, the Maccura F880 provides a balance of automation, accuracy, and diagnostic depth, making it a powerful tool for clinical hematology laboratories.
Category | Details |
---|---|
Detection Technology | Nucleic Acid Fluorescence Staining & Flow Cytometry for WBC, NRBC; 6-part Impedance Method and Flow Cytometry for RBC/PLT |
Detection Modes | CBC, DIFF, NRBC, RET, PLTF, AWS, SR |
Sample Modes | Whole Blood Mode, Low Value Leukocyte Mode, Predilution Mode, Sample Research Mode |
Sample Volume | Whole Blood Mode: 88 µL; Predilution Mode: 70 µL |
Throughput |
CBC+DIFF: 100 Tests/hour; CBC+DIFF+RET: 83 Tests/hour CBC+DIFF+RET+PLTF: 47 Tests/hour CBC+DIFF+RET+AWS: 71 Tests/hour CBC+DIFF+RET+PLTF+AWS: 47 Tests/hour |
Reporting Parameters (37 in total) |
Leukocyte: WBC, NEUT (#, %), LYMPH (#, %), MONO (#, %), EO (#, %), BASO (#, %), IG (#, %) Erythrocyte: RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, NRBC (#, %) Platelets: PLT, PDW, MPV, P-LCR, P-LCC, PCT, IPF Reticulocytes: RET (#, %), IRF, LFR, MFR, HFR, RET-He |
Auto Loader | Up to 50 sample positions |
Linear Range |
WBC: 0–500 × 10⁹/L RBC: 0–8.60 × 10¹²/L HGB: 0–260 g/L PLT: 0–5000 × 10⁹/L |